To the top of the document
Orlando
   
GMDE Start Page Load static TOC Load dynamic TOC Help?

Testing for Continuity

Special Tools

EL-39200 Digital Multimeter (DMM)

For equivalent regional tools, refer to Special Tools .

Caution: Refer to Test Probe Caution in the Preface section.

The following procedures verify good continuity in a circuit.

With a Digital Multimeter (DMM)

  1. Set the DMM to the Ω position.
  2. Disconnect the power feed (i.e. fuse, control module) from the suspect circuit.
  3. Disconnect the load.
  4. Select the MIN MAX function on the DMM.
  5. Connect one lead of the DMM to one end of the circuit to be tested.
  6. Connect the other lead of the DMM to the other end of the circuit.
  7. If the DMM displays low or no resistance, the circuit has good continuity.

With a Test Lamp

Note: Only use the test lamp procedure on low impedance power and ground circuits.

  1. Remove the power feed (i.e. fuse, control module) from the suspect circuit.
  2. Disconnect the load.
  3. Connect one lead of the test lamp to one end of the circuit to be tested.
  4. Connect the other lead of the test lamp to battery positive voltage.
  5. Connect the other end of the circuit to ground.
  6. If the test lamp illuminates (full intensity), then the circuit has good continuity.
   


© Copyright Chevrolet. All rights reserved
   


© Copyright Chevrolet. All rights reserved